首页> 外文期刊>Applied optics >Comparison of nonintensified and intensified CCD detectors for laser-induced breakdown spectroscopy
【24h】

Comparison of nonintensified and intensified CCD detectors for laser-induced breakdown spectroscopy

机译:用于激光诱导击穿光谱的非增强型和增强型CCD检测器的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The performance and sensitivity of an intensified CCD array system and a nonintensified CCD array detector system are compared for laser-induced breakdown spectroscopy (LIBS). LIBS measurements were recorded in a calcium-based aerosol-seeded gas stream at ambient pressure. The signal-to-noise ratio based on the 393.37-nm calcium emission line was calculated as a function of detector delay with respect to the plasma-initiating laser pulse. Both ensemble-averaging and single-shot spectral analyses were performed. For all conditions, the intensified CCD system provided an enhanced signal-to-noise ratio compared with the nonintensified CCD system.
机译:比较了增强型CCD阵列系统和非增强型CCD阵列检测器系统的性能和灵敏度,以进行激光诱导击穿光谱(LIBS)。 LIBS测量记录在环境压力下的钙基气雾剂气流中。根据检测器相对于等离子体引发激光脉冲的延迟,计算出基于393.37 nm钙离子发射谱线的信噪比。进行了集合平均和单次光谱分析。在所有情况下,与未增强的CCD系统相比,增强的CCD系统提供了更高的信噪比。