首页> 外文期刊>Applied optics >Design and testing of a white-light, steady-state diffuse reflectance spectrometer for determination of optical properties of highly scattering systems
【24h】

Design and testing of a white-light, steady-state diffuse reflectance spectrometer for determination of optical properties of highly scattering systems

机译:用于确定高散射系统光学特性的白光稳态漫反射光谱仪的设计和测试

获取原文
获取原文并翻译 | 示例
           

摘要

We present a steady-state radially resolved diffuse reflectance spectrometer capable of measuring the absorption and transport scattering spectra of tissue-simulating phantoms over an adjustable 170-nm wavelength interval in the visible and near infrared. Measurements in a variety of phantoms are demonstrated over the relevant range of tissue optical properties, and the accuracy of the instrument is found to be approximately 10% in both scattering and absorption. Monte Carlo simulations designed to test the accuracy of the instrument are presented that support the experimental findings. (C) 1997 Optical Society of America
机译:我们提出了一种稳态径向分辨漫反射光谱仪,该光谱仪能够在可见光和近红外光的可调170 nm波长间隔内测量组织模拟体模的吸收和传输散射光谱。在组织光学特性的相关范围内,对各种体模进行了测量,结果表明,该仪器在散射和吸收方面的准确度均约为10%。提出了旨在测试仪器精度的蒙特卡洛模拟,以支持实验结果。 (C)1997年美国眼镜学会