【24h】

PET-CT scanner characterization for PET raw data use in biomedical research

机译:PET-CT扫描仪表征用于生物医学研究中的PET原始数据

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this paper is to describe the experiments and methods that led to the geometrical interpretation of new-generation commercial PET-CT scanners, finalized to off-line PET-based treatment verification in ion beam therapy. Typically, the geometrical correspondence between the image domain (i.e., the dicom PET) and the sinogram domain (i.e., the PET raw data) is not explicitly described by scanner vendors. Hence, the proposed characterization can be applied to commercial PET-CT scanners used in biomedical research, for the development of technologies and methods requiring the use of PET raw data, without having access to confidential information from the vendors.
机译:本文的目的是描述导致新一代商用PET-CT扫描仪进行几何解释的实验和方法,并最终确定为离子束治疗中基于离线PET的治疗验证。通常,扫描仪销售商没有明确描述图像域(即,dicom PET)和正弦图域(即,PET原始数据)之间的几何对应关系。因此,所提出的特征可以应用于生物医学研究中使用的商用PET-CT扫描仪,用于开发需要使用PET原始数据的技术和方法,而无需访问供应商的机密信息。