Patients with acute brain pathology requiring ferromagnetic bio-medical implants for on-going invasive monitoring are largely excluded from the benefits of MRI scanning. We evaluated the behaviour of a thermal diffusion cortical blood flow (TD-CBF) sensor both in vitro (phantom gelatin model) and in vivo environments in a high field strength MRI system.Two baboons underwent cranial subdural implantation of 2 TD-CBF sensors/hemisphere and a single left parietal sensor was implanted subcortically to determine any deleterious effects. Using standard MRI sequences, artefact size, thermal effects, current generation, movement and reliability of recordings were assessed during scanning.The deflection forces were negligible, no observable thermal effects were demonstrated, while wide fluctuations in cerebral blood flow recordings were recorded. Mean image artefact size for implanted sensors was 6 times larger than in vitro. Patients with an implanted TD-CBF sensor may be safely imaged provided the device is disconnected. The MRI images obtained are of an acceptable quality.
机译:患有急性脑部病理的患者需要铁磁生物医学植入物以进行持续的侵入性监测,而这些患者在很大程度上被排除在MRI扫描之外。我们在高场强MRI系统中评估了热扩散皮层血流(TD-CBF)传感器在体外(幻影明胶模型)和体内环境的行为。两个狒狒进行了颅下硬膜下植入2枚TD-CBF传感器/皮下植入半球和单个左顶壁传感器,以确定是否有有害作用。使用标准的MRI序列,在扫描过程中评估了伪像的大小,热效应,电流的产生,运动和可靠性。偏转力可忽略不计,未观察到热效应,同时记录了脑血流记录的广泛波动。植入传感器的平均图像伪像大小是体外的6倍。如果设备断开连接,则可以对植入了TD-CBF传感器的患者进行安全的成像。获得的MRI图像质量合格。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号