The higher than expected early failure rate of the Sprint Fidelis implantable cardioverter defibrillator lead (model 6949, Medtronic, Inc.).

机译:Sprint Fidelis植入式心脏复律除颤器导线(型号6949,Medtronic,Inc。)的早期故障率高于预期。

获取原文
获取原文并翻译 | 示例
           

摘要

In the July 2007 print edition (online April 2007) of the Journal, Hauser et al reported the higher than expected early failure rate of the Sprint Fidelis implantable cardio-verter defibrillator lead (model 6949, Medtronic, Inc.) and recommended limiting its use until the issue was resolved.Over the last two and a half years, 180 Sprint Fidelis model 6931 (single coil) leads have been implanted and 5 have failed after an average of 16 months (range 13-19 months). In one case the lead malfunction was detected on routine device interrogation (impedance rise and electrical noise on the pace/sense channel).
机译:在《日刊》的2007年7月印刷版(2007年4月在线)中,Hauser等人报道了Sprint Fidelis植入式心脏除颤器导线(6949型,Medtronic,Inc.)的早期失败率高于预期,并建议限制使用在过去的两年半中,平均平均16个月(13-19个月)后,已植入180条Sprint Fidelis 6931型(单线圈)引线,其中5条失败。在一种情况下,在常规设备询问中检测到铅故障(阻抗升高和步速/感测通道上的电噪声)。