Extracorporeal pressure monitoring and the detection of vascular access stenosis.

机译:体外压力监测和血管通路狭窄的检测。

获取原文
获取原文并翻译 | 示例
           

摘要

Prospective monitoring of static venous pressure is an established tool to detect outflow stenoses in a vascular access. However, with this method it is not possible to identify vascular stenoses which are localized between the arterial and venous dialysis needle. We describe a new approach based on both static arterial and venous extracorporeal pressures. Pressure data of 9 dialysis patients with normal vascular access function and 9 patients with stenotic access were analyzed. Extracorporeal pressure was found to depend on the position of the heart relative to the extracorporeal blood circuit. All patients with venous outflow stenoses had an elevated ratio of arterial and venous intra-access pressure to mean arterial pressure. In case of access stenosis between arterial and venous needle the ratio of venous pressure to mean arterial pressure was normal, and only the arterial pressure ratio was elevated. We conclude that combined arterial and venous intraaccess pressure measurement normalized by mean blood pressure detects venous stenosis as well as stenosis between the arterial and venous dialysis needle. To minimize the rate of access thrombosis both arterial and venous intra-access pressure should be monitored.
机译:静静脉压的前瞻性监测是检测血管通路中流出狭窄的成熟工具。但是,用这种方法不可能识别位于动脉和静脉透析针之间的血管狭窄。我们描述了一种基于静态动脉和静脉体外压力的新方法。分析了9例血管通路正常的透析患者和9例狭窄通路的患者的压力数据。发现体外压力取决于心脏相对于体外血液循环的位置。所有患有静脉流出狭窄的患者的动脉和静脉内进入压力与平均动脉压力之比均升高。在动脉和静脉针之间进入狭窄的情况下,静脉压力与平均动脉压力之比是正常的,只有动脉压力比升高。我们得出的结论是,通过平均血压归一化的联合动脉和静脉内通路压力测量可检测静脉狭窄以及动脉和静脉透析针之间的狭窄。为了使通路血栓形成的速率最小化,应同时监测动脉和静脉的通路内压力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号