Assessment of polyethylene stability using the standardized OIT procedure by DSC method

机译:通过DSC方法使用标准化OIT程序评估聚乙烯的稳定性

获取原文
获取原文并翻译 | 示例
           

摘要

The standardized procedure for determination of oxidation induction time(OIT)using the differential scanning calorimetry(DSC)method has been described.A Perkin-Elmer DSC7 instrument was used and it was calibrated with indium and tin certified reference materials(CRMs).Measured values of melting temperature were consistent with the certified values within ±0.3 °C.The procedure was validated using commercially available PE samples.Relative standard deviation(RSD)values were calculated and compared with the published data.The values obtained were lower than those quoted in literature for particular PE grades.Then chemically modified(grafted)PE matrices and composites with a flame retardant filler were investigated.OIT values were measured at various temperatures and the optimum measurement temperature for modified PE composites was determined.It was equal to 220 deg C as compared with 200 deg C for unmodified PE.Thus,it has been found that the thermal stability of investigated composites in thermooxi-dative degradation conditions is higher than that of unmodified PE material.
机译:描述了使用差示扫描量热法(DSC)测定氧化诱导时间(OIT)的标准化程序。使用Perkin-Elmer DSC7仪器,并用铟和锡认证的参考物质(CRM)进行校准。熔点温度与±0.3°C范围内的认证值相符。使用市售PE样品验证了该程序。计算了相对标准偏差(RSD)值并将其与公开数据进行了比较。针对特定PE等级的文献进行了研究,然后研究了化学改性的(接枝)PE基质和含阻燃填料的复合材料,在各种温度下测量了OIT值,并确定了改性PE复合材料的最佳测量温度,等于220摄氏度与未改性PE的200摄氏度相比。因此,已发现所研究复合材料的热稳定性在热氧化降解条件下高于未改性的PE材料。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号