Evaluation of accuracy of a theoretical model for predicting thenecrosed tissue volume during focused ultrasound surgery

机译:评估超声聚焦手术中坏死组织体积的理论模型的准确性评估

获取原文
获取原文并翻译 | 示例
           

摘要

The concept of thermal dose as a predictor for the size of thennecrosed tissue volume during high-intensity focussed ultrasound surgerynwas tested. The sensitivity of the predicted lesion size to thenuncertainties in the iso-dose constant, attenuation coefficient, andnthermal dose threshold of necrosis was studied. The predicted lesionnsize appears to be independent of attenuation at some high attenuationnvalues and certain depth in tissue. Thus, for a given target depth, anproper selection of frequency could minimize the lesion size variabilityndue to uncertainty in the tissue attenuation. The predicted lesion sizenwas less dependent on the uncertainties in the iso-dose constant andnthermal dose of necrosis. The predictions of the model were comparednwith experimental data in rabbit muscle, and experimental data in catnand rat brain measured by others. The agreement was found to be good innmost of the experiments. Similarly, the model was found to predict wellnthe trends of increasing power and pulse duration
机译:测试了热剂量作为高强度聚焦超声手术中坏死组织体积大小的预测指标的概念。研究了在坏死的等剂量常数,衰减系数和非热剂量阈值下,预测病变大小对不确定度的敏感性。预测的病变大小似乎与某些高衰减值和组织中一定深度处的衰减无关。因此,对于给定的目标深度,由于组织衰减的不确定性,适当选择频率可以使病变大小的可变性最小。预计的病变大小较少依赖于坏死等剂量恒定和非热剂量的不确定性。将该模型的预测结果与兔肌肉的实验数据进行比较,并通过其他方法对模型的脑和大鼠脑中的实验数据进行比较。发现该协议在大多数实验中都是很好的。同样,发现该模型可以预测功率和脉冲持续时间增加的趋势

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号