Ultrasound Transducer and System for Real-Time Simultaneous Therapy and Diagnosis for Noninvasive Surgery of Prostate Tissue

机译:超声换能器和前列腺组织无创手术实时同时治疗和诊断的系统

获取原文
获取原文并翻译 | 示例
           

摘要

For noninvasive treatment of prostate tissue using high-intensity focused ultrasound this paper proposes a design of an integrated multifunctional confocal phased array (IMCPA) and a strategy to perform both imaging and therapy simultaneously with this array. IMCPA is composed of triplerow phased arrays: a 6-MHz array in the center row for imaging and two 4-MHz arrays in the outer rows for therapy. Different types of piezoelectric materials and stack configurations may be employed to maximize their respective functionalities, i.e., therapy and imaging. Fabrication complexity of IMCPA may be reduced by assembling already constructed arrays. In IMCPA, reflected therapeutic signals may corrupt the quality of imaging signals received by the center-row array. This problem can be overcome by implementing a coded excitation approach and/or a notch filter when B-mode images are formed during therapy. The 13-bit Barker code, which is a binary code with unique autocorrelation properties, is preferred for implementing coded excitation, although other codes may also be used. From both Field II simulation and experimental results, we verified whether these remedial approaches would make it feasible to simultaneously carry out imaging and therapy by IMCPA. The results showed that the 13-bit Barker code with 3 cycles per bit provided acceptable performances. The measured -6 dB and -20 dB range mainlobe widths were 0.52 mm and 0.91 mm, respectively, and a range sidelobe level was measured to be -48 dB regardless of whether a notch filter was used. The 13-bit Barker code with 2 cycles per bit yielded -6 dB and -20 dB range mainlobe widths of 0.39 mm and 0.67 mm. Its range sidelobe level was found to be -40 dB after notch filtering. These results indicate the feasibility of the proposed transducer design and system for real-time imaging during therapy.
机译:对于使用高强度聚焦超声进行前列腺组织的非侵入性治疗,本文提出了一种集成多功能共聚焦相控阵(IMCPA)的设计,以及采用该阵列同时进行成像和治疗的策略。 IMCPA由三行相控阵组成:在中心行有一个6 MHz的阵列进行成像,在外部行有两个4 MHz的阵列进行治疗。可以采用不同类型的压电材料和叠层结构来最大化其各自的功能,即治疗和成像。通过组装已经构造的阵列,可以降低IMCPA的制造复杂性。在IMCPA中,反射的治疗信号可能会破坏中心行阵列接收到的成像信号的质量。当在治疗过程中形成B模式图像时,可以通过实施编码激励方法和/或陷波滤波器来克服此问题。尽管还可以使用其他代码,但是13位的巴克代码是具有唯一自相关属性的二进制代码,是实现编码激励的首选方法。从Field II模拟和实验结果中,我们验证了这些补救方法是否使同时进行IMCPA成像和治疗可行。结果表明,每位3个周期的13位Barker代码提供了可接受的性能。测得的-6 dB和-20 dB范围主瓣宽度分别为0.52 mm和0.91 mm,并且无论是否使用陷波滤波器,测得的范围旁瓣电平均为-48 dB。每位2个周期的13位Barker码产生的-6 dB和-20 dB范围主瓣宽度分别为0.39 mm和0.67 mm。陷波滤波后,其范围旁瓣电平为-40 dB。这些结果表明提出的换能器设计和系统在治疗过程中实时成像的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号