首页> 外文期刊>Journal of Clinical Microbiology >Growth Detection Failures by the Nonradiometric Bactec MGIT 960 Mycobacterial Culture System
【24h】

Growth Detection Failures by the Nonradiometric Bactec MGIT 960 Mycobacterial Culture System

机译:非辐射Bactec MGIT 960分枝杆菌培养系统的生长检测失败

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Mycobacterial growth in liquid culture can go undetected by automated, nonradiometric growth detection systems. In our laboratory, instrument-negative tubes from the Bactec MGIT 960 system are inspected visually for clumps suggestive of mycobacterial growth, which (if present) are examined by acid-fast smear analysis. A 3-year review demonstrated that ~1% of instrument-negative MGIT cultures contained mycobacterial growth and that 10% of all cultures yielding mycobacteria were instrument negative. Isolates from instrument-negative MGIT cultures included both tuberculous and nontuberculous mycobacteria.
机译:自动化的非放射线生长检测系统无法检测到液体培养中的分枝杆菌生长。在我们的实验室中,对Bactec MGIT 960系统的仪器阴性试管进行目视检查,以发现提示分枝杆菌生长的团块,并通过耐酸涂片分析对其进行检查。历时3年的审查证明,约有1%的器械阴性MGIT培养物含有分枝杆菌生长,而所有产生分枝杆菌的培养物中有10%呈器械阴性。器械阴性MGIT培养物的分离物包括结核和非结核分枝杆菌。