Comparison of BacT/Alert with Signal blood culture system.

机译:BacT / Alert与Signal血液培养系统的比较。

获取原文
           

摘要

The BacT/Alert (Organon Teknika Corp., Durham, N.C.) is an automated blood culture system. It is based on the detection of CO2 by means of a colorimetric sensor internally attached to the bottom of culture bottles. The aerobic and anaerobic media of this system were compared with one bottle of the Signal system (Oxoid Ltd., Hampshire, United Kingdom). At bedside, 20 ml of blood was drawn from each adult patient. The two BacT/Alert bottles were inoculated with 5 ml of blood each; the Signal bottle was inoculated with 10 ml. A total of 5,284 sets (2,483 patients; 2.1 cultures per patient) consisting of three bottles each were evaluated, of which 781 sets (14.8%) revealed microorganisms (n = 892); 642 of these were considered to be pathogenic. Significantly more (P < 0.0001) pathogens were isolated from the two BacT/Alert bottles together (n = 584) than from the single Signal bottle (n = 515). Escherichia coli (P = 0.007), gram-negative bacteria other than members of the family Enterobacteriaceae or Pseudomonas spp. (P = 0.006), and yeasts (P = 0.02) were isolated more often from both or either BacT/Alert bottle. Comparing the systems in terms of 388 different organisms per septic episode, the difference between BacT/Alert and Signal was significant for the total number of septicemia cases (P = 0.003). More contaminants grew in the BacT/Alert system (173 versus 116; P = 0.0001). False-positive indications were more frequent in the BacT/Alert system, 198 (3.7%) aerobic bottles and 57 (1.1%) anaerobic bottles, than in the Signal bottles, 24 (0.5%) bottles. Pathogens could be detected significantly earlier (P < 0.0001) in the BacT/Alert system than in the Signal system. The BacT/Alert instrument with two bottles allowed earlier detection as well as the isolation of more microorganisms than the manual, one-bottle Signal system.
机译:BacT / Alert(Organon Teknika Corp.,Durham,N.C.)是一种自动血液培养系统。它基于内部连接到培养瓶底部的比色传感器检测到的CO2。将该系统的需氧和厌氧培养基与一瓶Signal系统(英国汉普郡的Oxoid公司)进行了比较。在床边,从每位成年患者中抽取20毫升血液。向两个BacT / Alert瓶分别接种5毫升血液;用10 ml接种信号瓶。共评估了5284套(2483名患者;每名患者2.1种培养物),每套三瓶,其中781套(14.8%)的微生物被发现(892例)。其中642个被认为是致病的。从两个BacT / Alert瓶(n = 584)中分离出的病原体明显多于从单个Signal瓶(n = 515)中分离出的病原体(P <0.0001)。大肠杆菌(P = 0.007),除肠杆菌科或假单胞菌属之外的革兰氏阴性细菌。 (P = 0.006),并且从BacT / Alert瓶或BacT / Alert瓶中分离酵母的频率更高(P = 0.02)。比较每个败血症事件中388种不同生物体的系统,对于败血症病例总数,BacT / Alert和Signal之间的差异是显着的(P = 0.003)。 BacT / Alert系统中的污染物增多(173对116; P = 0.0001)。在BacT / Alert系统中,有198个(3.7%)需氧瓶和57个(1.1%)厌氧瓶比在Signal瓶中有24个(0.5%)误报率更高。在BacT / Alert系统中,与在信号系统中相比,病原体的检测时间要早得多(P <0.0001)。与手动一瓶信号系统相比,带有两个瓶子的BacT / Alert仪器可以更早地进行检测并分离出更多的微生物。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号