...
首页> 外文期刊>Journal of Clinical Microbiology >Comparison of the BacT/Alert PF Pediatric FAN Blood Culture Bottle with the Standard Pediatric Blood Culture Bottle, the Pedi-BacT
【24h】

Comparison of the BacT/Alert PF Pediatric FAN Blood Culture Bottle with the Standard Pediatric Blood Culture Bottle, the Pedi-BacT

机译:BacT / Alert PF小儿FAN血培养瓶与标准小儿血培养瓶Pedi-BacT的比较

获取原文
           

摘要

The performance of the BacT/Alert PF (Organon-Teknika Corp., Durham, N.C.), a new nonvented pediatric FAN blood culture bottle, was compared to that of the original pediatric bottle, the Pedi-BacT, with matched aerobic cultures obtained from two separate facilities. A total of 244 clinically significant isolates were recovered from 4,015 compliant pairs. Among the positive cultures, 170 (70%) isolates were detected in both the BacT/Alert PF and the Pedi-BacT bottles, while 47 (19%) isolates were recovered in the BacT/Alert PF bottle only and 27 (11%) isolates were recovered in the Pedi-BacT bottle only. Although isolation of specific microorganisms was comparable for the two bottles, the total number of organisms recovered by the BacT/Alert PF was greater than that by the Pedi-BacT (P = 0.0272). In addition, more organisms were recovered by the BacT/Alert PF bottle from the blood of patients receiving antimicrobial therapy (P = 0.0180). Overall time to detection was similar for the two bottles; however, a significantly decreased mean time to detection was recorded for yeast from the BacT/Alert PF bottle (22.9 h; P = 0.0001) and staphylococci from the Pedi-BacT bottle (22.5 h; P = 0.0056). One false-negative culture and five false-positive cultures occurred with the Pedi-BacT bottle, compared to one false-positive culture with the BacT/Alert PF bottle. The BacT/Alert PF bottle is a reliable blood culture bottle for pediatric blood culture specimens and may offer improved recovery of microbes from patients on antimicrobial therapy. The use of the nonvented bottle will both facilitate bottle processing and decrease expenditures for materials due to the elimination of the venting needles required for the original vented bottles.
机译:将BacT / Alert PF(北卡罗莱纳州达勒姆市的Organon-Teknika公司)的性能与新型无通风FAN血液培养瓶的性能进行了比较,将其与原始的儿科瓶Pedi-BacT进行了比较,并获得了匹配的需氧培养两个独立的设施。从4,015对顺应性对中共回收了244种临床上重要的分离株。在阳性培养物中,在BacT / Alert PF和Pedi-BacT瓶中均检测到170种(70%)分离株,而仅在BacT / Alert PF瓶中回收了47(19%)株,而27(11%)分离株仅在Pedi-BacT瓶中回收。尽管在这两个瓶子中分离出的特定微生物相当,但BacT / Alert PF回收的生物总数大于Pedi-BacT( P = 0.0272)。此外,通过BacT / Alert PF瓶从接受抗微生物治疗的患者的血液中回收了更多的生物( P = 0.0180)。这两个瓶子的总检测时间相似。但是,BacT / Alert PF瓶中的酵母(22.9 h; P = 0.0001)和Pedi-BacT瓶中的葡萄球菌(22.5 h; P = 0.0056)。与使用BacT / Alert PF瓶的一种假阳性培养物相比,Pedi-BacT瓶发生了一种假阴性培养物和五种假阳性培养物。 BacT / Alert PF瓶是用于小儿血液培养标本的可靠血液培养瓶,可以提高接受抗菌治疗的患者微生物的回收率。由于省去了原始排气瓶所需的排气针,因此使用非排气瓶既可以促进瓶的加工,又可以减少材料的支出。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号