Use of the Autobac system for detection of methicillin-resistant Staphylococcus aureus.

机译:使用Autobac系统检测耐甲氧西林的金黄色葡萄球菌。

获取原文
           

摘要

Using 490 strains of Staphylococcus aureus divided into methicillin-susceptible, -resistant, and -heteroresistant varieties, we compared the results obtained by the agar disk method with those obtained with the automated Autobac system. Susceptible strains exhibited a perfect correlation, whereas there were numerous discrepancies with resistant and still more with heteroresistant varieties. When incubation was increased to 18 h at 37 degrees C (Autobac incubation temperature), 35 degrees C, or 30 degrees C, these differences disappeared, but other problems may arise when incubation is prolonged, especially with erythromycin. We thus recommend carrying out two readings, a normal one after 3 h of incubation and a special reading after 18 h, solely for the detection of heteroresistance to methicillin.
机译:使用490株金黄色葡萄球菌,分为对甲氧西林敏感,耐药和对异菌的变种,我们比较了琼脂圆盘法与自动Autobac系统获得的结果。易感菌株表现出完美的相关性,而抗病品种之间存在许多差异,而异种抗性品种之间的差异更大。当在37摄氏度(Autobac孵育温度),35摄氏度或30摄氏度的条件下将孵育时间增加到18小时时,这些差异消失了,但是如果延长孵育时间(尤其是红霉素),则可能会出现其他问题。因此,我们建议进行两次读数,孵育3小时后正常读数,而18小时后进行特殊读数,仅用于检测对甲氧西林的异抗性。