The feasibility of measuring dose distributions in three-dimensions using Gafchromic EBT film was studied. A stack of film was used as tomographic dosimeter. The response of film (4×4 cm2) placed between a stack of Gafchromic EBT films and slabs of polystyrene phantom was tested. Differences in film response for two different photon beam qualities, 6 MV and 25 MV, were observed. Measurements with 6 MV beam quality revealed that for different field sizes and depths there were no changes in response of EBT Gafchromic film when placed between polystyrene plates or sandwiched between other films. However, in the case of 25 MV beam quality, a statistically significant over-response was found for film placed in stack. We conclude that stacked-film dosimetry is feasible at 6 MV. For 25 MV photon beam quality the stack of films showed different properties than the polystyrene slabs of the same height.
机译:研究了使用Gafchromic EBT薄膜测量三维剂量分布的可行性。一堆薄膜用作断层摄影剂量计。测试膜(4×4cm2)置于堆叠在一堆Gafchromic EBT薄膜和聚苯乙烯体映像板之间的响应。观察到两种不同光子束质量,6mV和25mV的膜响应的差异。具有6个MV光束质量的测量显示,对于不同的场尺寸和深度,在聚苯乙烯板或夹在其他薄膜之间时,EBT Gafchromic膜的响应没有变化。然而,在25 mV光束质量的情况下,发现堆叠在堆叠中的薄膜具有统计学上显着的过响应。我们得出结论,堆叠薄膜剂量测定法在6mV下可行。对于25 MV光子光束质量,薄膜叠叠显示不同的性质比相同高度的聚苯乙烯片。