【24h】

Novel CO2 laser system for hard tissue ablation

机译:用于硬组织消融的新型CO2激光系统

获取原文
获取原文并翻译 | 示例

摘要

Abstract: A line-tunable and variable pulselength CO$-2$/ laser system for ablation of dental and bone tissues has been tested. To avoid the undesirable effects of plasma formation, acoustic shock waves and collateral thermal damage to the tissues, we have found an optimum laser pulselength in the 10 - 20 microseconds range. At a fluence of 30 J/cm$+2$/ and laser wavelength of 10.6 $mu@m the ablation depths in enamel (dentin) were 12(18) $mu@m and no significant noise was observed. Without water cooling enamel showed no visible color change, while some minimal carbonization of dentin was present even at low pulse repetition frequencies. !10
机译:摘要:已经测试了用于消融牙齿和骨组织的线可调和可变脉冲长度CO $ -2 $ /激光系统。为避免等离子体形成,声波冲击和对组织的附带热损害的不良影响,我们发现最佳激光脉冲长度在10-20微秒的范围内。在注量为30 J / cm $ + 2 $ /和激光波长为10.6μμm时,牙釉质(牙质)的消融深度为12(18)μμm,未观察到明显的噪声。没有水冷却,牙釉质没有可见的颜色变化,而即使在低的脉冲重复频率下,牙本质也有一些最小的碳化。 !10