AN EVALUATION OF THE STRYKER 90D PEDICLE SCREW FOR USE AS A REVISION SCREW

机译:用于修订版的STRYKER 90D脚踏螺丝的评估

获取原文
获取原文并翻译 | 示例

摘要

The results of this work indicate that the Stryker 90D pedicle screw is a good screw design for use as a revision screw for a wide range of pedicle screw designs. The 90D is within 15% of the strength of the pullout strength of the original screw for all but two screw designs tested. Of the seventeen screw designs tested the two designs in which the Stryker 90D performed poorly as a revision screw were for the Omega 21 and the XIA Titanium.
机译:这项工作的结果表明,Stryker 90D椎弓根螺钉是一种很好的螺钉设计,可以用作各种椎弓根螺钉设计的修正螺钉。对于90D,除了两个测试的螺钉设计外,其余所有螺钉的强度均在原始螺钉的拉出强度的15%以内。在测试的17种螺钉设计中,有2种设计的Stryker 90D表现不佳,其中Omega 21和XIA Titanium的改正螺钉表现不佳。