Increasing numbers of devices are emerging which involve biophotonic imaging on a mobile platform. Therefore, effective test methods are needed to ensure that these devices provide a high level of image quality. We have developed novel phantoms for performance assessment of near infrared fluorescence (NIRF) imaging devices. Resin molding and 3D printing techniques were applied for phantom fabrication. Comparisons between two imaging approaches - a CCD-based scientific camera and an NIR-enabled mobile phone - were made based on evaluation of the contrast transfer function and penetration depth. Optical properties of the phantoms were evaluated, including absorption and scattering spectra and fluorescence excitation-emission matrices. The potential viability of contrast-enhanced biological NIRF imaging with a mobile phone is demonstrated, and color-channel-specific variations in image quality are documented. Our results provide evidence of the utility of novel phantom-based test methods for quantifying image quality in emerging NIRF devices.
机译:出现越来越多的涉及在移动平台上进行生物光子成像的设备。因此,需要有效的测试方法来确保这些设备提供高水平的图像质量。我们已经开发出用于近红外荧光(NIRF)成像设备性能评估的新型体模。树脂成型和3D打印技术被应用于幻影制造。基于对比传递函数和穿透深度的评估,对两种成像方法(基于CCD的科学相机和支持NIR的手机)进行了比较。评价了体模的光学性质,包括吸收和散射光谱以及荧光激发-发射矩阵。演示了使用手机进行对比度增强的生物NIRF成像的潜在可行性,并记录了图像质量中特定于颜色通道的变化。我们的结果提供了新颖的基于幻像的测试方法来量化新兴NIRF设备中图像质量的效用的证据。

著录项

  • 来源
  • 会议地点 San Francisco CA(US)
  • 作者单位

    University of Maryland, Dept. of Bioengineering, College Park, USA,Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD USA 20993;

    Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD USA 20993;

    Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD USA 20993;

    Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD USA 20993;

    University of Maryland, Dept. of Bioengineering, College Park, USA;

    Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD USA 20993;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Near infrared fluorescence; Performance testing; Mobile phone; 3D printing; Tissue phantoms;

    机译:近红外荧光;性能测试;移动电话; 3D打印组织体模;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号