The accuracy and precision of radiostereometric analysis in upper limb arthroplasty

机译:上肢置换术中放射立体分析的准确性和准确性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background and purpose - Radiostereometric analysis (RSA) is an accurate method for measurement of early migration of implants. Since a relation has been shown between early migration and future loosening of total knee and hip prostheses, RSA plays an important role in the development and evaluation of prostheses. However, there have been few RSA studies of the upper limb, and the value of RSA of the upper limb is not yet clear. We therefore performed a systematic review to investigate the accuracy and precision of RSA of the upper limb. Patients and methods - PRISMA guidelines were followed and the protocol for this review was published online at PROSPERO under registration number CRD42016042014. A systematic search of the literature was performed in the databases Embase, Medline, Cochrane, Web of Science, Scopus, Cinahl, and Google Scholar on April 25, 2015 based on the keywords radiostereometric analysis, shoulder prosthesis, elbow prosthesis, wrist prosthesis, trapeziometacarpal joint prosthesis, humerus, ulna, radius, carpus. Articles concerning RSA for the analysis of early migration of prostheses of the upper limb were included. Quality assessment was performed using the MINORS score, Downs and Black checklist, and the ISO RSA Results - 23 studies were included. Precision values were in the 0.06-0.88 mm and 0.05-10.7° range for the shoulder, the 0.05-0.34 mm and 0.16-0.76° range for the elbow, and the 0.16-1.83 mm and 11-124° range for the TMC joint. Accuracy data from marker- and model-based RSA were not reported in the studies included. Interpretation - RSA is a highly precise method for measurement of early migration of orthopedic implants in the upper limb. However, the precision of rotation measurement is poor in some components. Challenges with RSA in the upper limb include the symmetrical shape of prostheses and the limited size of surrounding bone, leading to over-projection of the markers by the prosthesis. We recommend higher adherence to RSA guidelines and encourage investigators to publish long-term follow-up RSA studies
机译:背景与目的-放射立体分析(RSA)是一种用于测量植入物早期迁移的准确方法。由于已经显示出早期迁移与将来整个膝关节和髋关节假体松动之间的关系,因此RSA在假体的开发和评估中起着重要的作用。然而,关于上肢的RSA研究很少,并且上肢的RSA的价值还不清楚。因此,我们进行了系统的审查,以调查上肢RSA的准确性和准确性。患者和方法-遵循PRISMA指南,本评价的方案在线发布于PROSPERO,注册号为CRD42016042014。系统在2015年4月25日对Embase,Medline,Cochrane,Web of Science,Scopus,Cinahl和Google Scholar等数据库进行了系统的文献检索,这些关键词基于放射性立体分析,肩关节假体,肘关节假体,腕关节假体,梯形腕骨关节假体,肱骨,尺骨,radius骨,腕骨。包括有关RSA的文章,用于分析上肢假体的早期迁移。使用MINORS分数,Downs和Black核对表以及ISO RSA结果进行了质量评估-包括23项研究。肩膀的精度值在0.06-0.88 mm和0.05-10.7°范围内,肘部的精度在0.05-0.34 mm和0.16-0.76°范围内,而TMC关节的精度在0.16-1.83 mm和11-124°范围内。包括在研究中没有报告基于标记和基于模型的RSA的准确性数据。解释-RSA是一种高度精确的方法,用于测量上肢骨科植入物的早期迁移。但是,某些组件的旋转测量精度较差。上肢RSA面临的挑战包括假体的对称形状和周围骨骼的有限大小,从而导致假体过度投射标记物。我们建议您更好地遵守RSA准则,并鼓励研究者发布长期随访RSA研究

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号