>The reliability of continuous noninvasive finger blood pressure measurement in critically ill children
【2h】
The reliability of continuous noninvasive finger blood pressure measurement in critically ill children
展开▼
机译:危重儿童连续无创手指血压测量的可靠性
展开▼
免费
页面导航
摘要
著录项
引文网络
相似文献
相关主题
摘要
INTRODUCTION: Continuous noninvasive arterial blood pressure can be measured in finger arteries using an inflatable finger cuff (FINAP) with a special device and has proven to be feasible and reliable in adults. We studied prototype pediatric finger cuffs and pediatric software to compare this blood pressure measurement with intraarterially measured blood pressure (IAP) in critically ill children. METHODS: We included sedated and mechanically ventilated children admitted to our pediatric intensive care unit. We performed simultaneous arterial blood pressure measurements during a relatively stable hemodynamic period and compared FINAP, IAP, and the noninvasive blood pressure oscillometric technique. We also compared IAP to a reconstruction of brachial pressure from finger pressure. RESULTS: Thirty-five children between 2 and 22 kg body weight were included. In total, 152 attempts to record a FINAP pressure were performed of which 4.6% were unsuccessful. When comparing FINAP to IAP, bias was -16.2, -7.7, and -10.2 mm Hg for systolic arterial blood pressure, diastolic arterial blood pressure, and mean arterial blood pressure. Limits of agreement (LOA) were respectively 26.1%, 30.1%, and 22.6%. When reconstruction of brachial pressure from finger pressure was compared to IAP, these results were -11.8, 0.6, and -0.9 mm Hg for bias and 21.7%, 8.9%, and 8.9% for LOA. When noninvasive blood pressure oscillometric technique was compared to IAP, the results were: -6.8, -0.9, and -3.8 mm Hg for bias and 18.2%, 38.6%, and 22.1% for LOA. CONCLUSION: Beta type continuous noninvasive arterial blood pressure monitoring using a finger cuff with brachial arterial waveform reconstruction seems reliable in hemodynamically stable critically ill children
展开▼
机译:简介:可以使用带有特殊装置的充气式指套(FINAP)在手指动脉中测量连续的非侵入性动脉血压,并已证明在成人中是可行和可靠的。我们研究了原型儿科手指套囊和儿科软件,以比较该危重症患儿的血压测量值与动脉内测量血压(IAP)的差异。方法:我们纳入了入院小儿重症监护室的镇静和机械通气儿童。我们在相对稳定的血液动力学期间进行了动脉血压测量,并比较了FINAP,IAP和无创血压示波法。我们还将IAP与通过手指压力重建肱动脉压力进行了比较。结果:包括35名体重在2到22公斤之间的儿童。总共进行了152次尝试记录FINAP压力的尝试,其中4.6%未成功。比较FINAP和IAP时,收缩压,舒张压和平均动脉压的偏倚分别为-16.2,-7.7和-10.2 mm Hg。协议限制(LOA)分别为26.1%,30.1%和22.6%。将手指压力重建肱动脉压力与IAP进行比较时,偏见的结果为-11.8、0.6和-0.9 mm Hg,LOA的结果为21.7%,8.9%和8.9%。将无创血压示波法与IAP进行比较时,偏倚的结果为:-6.8,-0.9和-3.8 mm Hg,LOA的结果为18.2%,38.6%和22.1%。结论:使用手指袖套与肱动脉波形重建术进行β型连续无创血压监测在血流动力学稳定的危重儿童中似乎是可靠的
展开▼