Various calibration procedures result in optimal standardization of routinely used 25(OH)D ID-LC-MS/MS methods

机译:各种校准程序可以使常规使用的25(OH)D ID-LC-MS / MS方法达到最佳标准化

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: The variety of LC-MS/MS methods measuring total 25(OH)D used today is vast and the comparability among these methods is still not well assessed. Methods: Here, we performed a comparison in samples of healthy donors between the currently routinely used 25(OH)D LC-MS/MS methods in the Netherlands and the Ghent University reference measurement procedure to address this issue (n = 40). Additionally, an interlaboratory comparison in patient serum samples assessed agreement between the Dutch diagnostic methods (n = 37). Results: The overall correlation of the routine methods for 25(OH)D-3 with the reference measurement procedures and with the mean of all diagnostic methods was excellent (r > 0.993 and r > 0.989, respectively). Three out of five methods aligned perfectly with both the reference measurement procedure and the median of all methods. One of the routine methods showed a small positive bias, while another showed a small negative bias consistently in both comparisons. Conclusion: The biases most probably originated from differences in calibration procedure and may be obviated by reassessing calibration of stock standards and/or calibrator matrices. In conclusion, five diagnostic centers have performed a comparison with the 25(OH)D Ghent University reference measurement procedure in healthy donor serum samples and a comparison among themselves in patient serum samples. Both analyses showed a high correlation and specificity of the routine LC-MS/MS methods, yet did reveal some small standardization issues that could not be traced back to the technical details of the different methods. Hence, this study indicates various calibration procedures can result in perfect alignment. (C) 2016 Elsevier B.V. All rights reserved.
机译:背景技术:目前,使用各种LC-MS / MS方法测量总25(OH)D的方法非常广泛,这些方法之间的可比性仍未得到很好的评估。方法:在这里,我们对健康的供体样本进行了比较,以比较荷兰目前常用的25(OH)D LC-MS / MS方法和根特大学的参考测量程序来解决此问题(n = 40)。此外,在患者血清样本中进行的实验室间比较评估了荷兰诊断方法之间的一致性(n = 37)。结果:25(OH)D-3常规方法与参考测量程序以及所有诊断方法的平均值之间的总体相关性极好(分别为r> 0.993和r> 0.989)。五分之三的方法与参考测量程序和所有方法的中位数均完美匹配。在两种比较中,一种常规方法显示出较小的正偏差,而另一种则始终显示出较小的负偏差。结论:偏差很可能源自校准程序的差异,并且可以通过重新评估库存标准品和/或校准剂基质的校准来消除。总之,五个诊断中心对健康供体血清样本中的25(OH)D根特大学参考测量程序进行了比较,并对患者血清样本中的它们进行了比较。两种分析均显示了常规LC-MS / MS方法的高度相关性和特异性,但确实揭示了一些小的标准化问题,这些问题无法追溯到不同方法的技术细节。因此,这项研究表明各种校准程序可以导致完美的对准。 (C)2016 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号