首页> 外文OA文献 >Methods development of multimodal loading under bending, compression and torsion for bench-top testing of cardiovascular stents
【2h】

Methods development of multimodal loading under bending, compression and torsion for bench-top testing of cardiovascular stents

机译:方法开发弯曲,压缩和扭转多模态负荷,用于心血管支架的台式试验

摘要

This thesis describes a method for setting up, testing and analyzing cardiovascular stents in multiple deformation modes. Testing stents in multiple deformation modes, including bending, torsion and compression simultaneously is a novel technique. Specific details are provided on strategies for inserting a stent sample into the testing apparatus such that stent integrity is preserved. Using a dynamic feedback cycle for buffered video capture we have been able to document sequential fracture events during testing scenarios. Stereomicroscopy further allows us to detect and determine placement of stent fracture. This method will help stent designers to create stents that are more likely to stand up to the particular conditions they will face in the body and regulatory agencies to determine how well potential new stents will fair under physiologic conditions.
机译:本文介绍了一种在多种变形模式下建立,测试和分析心血管支架的方法。在多种变形模式下同时测试支架,包括弯曲,扭转和压缩,是一种新颖的技术。提供了用于将支架样品插入测试设备中以保持支架完整性的策略的具体细节。通过使用动态反馈循环进行缓冲视频捕获,我们已经能够记录测试场景中的连续断裂事件。立体镜进一步使我们能够检测和确定支架骨折的位置。这种方法将帮助支架设计者制造出更能经受住人体和监管机构面对的特定条件的支架,从而确定潜在的新支架在生理条件下的适应能力。