Total-to-peak ratios of high purity germanium gamma ray detector

机译:高纯度锗伽马射线探测器的总峰比

摘要

This study is concerned with the percentage of [gamma]-rays of a certain energy having their energy correctly measured by a high purity Germanium [gamma]-ray detector. The ratio between the total counts and the counts within the energy peak (total-to-peak ratio) is determined for seven energies ranging from 89 keV to 1275 keV. A Monte Carlo based on the physical parameters of the detector was used to extrapolate between these points and after an energy independent scaling factor fit the data with a reduced [chi]² slightly below 1. The same experiment was repeated with a lead brick and then a β detector near the Ge detector and these objects were found to not have an effect on the total-to-peak ratios within the precision of the experiment.
机译:该研究涉及通过高纯度锗γ射线检测器正确测量其能量的一定能量的γ射线的百分比。确定了总计数与能量峰值内计数之间的比率(总峰比),确定的范围是89 keV至1275 keV的七个能量。使用基于检测器物理参数的蒙特卡洛法在这些点之间进行推断,并且在能量无关的比例因子拟合后,将数据的χ²降低至略低于1。然后使用铅砖重复相同的实验,然后在Ge探测器附近有一个β探测器,发现这些物体在实验精度范围内对总峰比没有影响。