A Disposable Polymer Lab-On-A-Slide For Point-Of-Care Diagnostics Of Methicillin-Resistant Staphylococcus Aureus (Mrsa)

机译:用于甲氧西林抗性金黄色葡萄球菌(mrsa)的即时诊断的一次性聚合物实验室-a-slide

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This paper reports the design, fabrication and experimental verification of a polymer microfluidic labon-a-slide for rapid detection of methicillin-resistant Staphylococcus aureus (MRSA). MRSA cells were captured in a lysis chamber using magnetic beads, followed by thermal lysis. The released DNA was transferred into a second chamber for polymerase chain reaction (PCR) amplification. Fluidic control in the device was accomplished by pneumatic actuation of a micropump and five microvalves integrated on the device. The mecA gene from MRSA was successfully amplified by real-time PCR within 35 min. Presence of the correct PCR amplicon (186 bp) was verified using capillary electrophoresis. The results demonstrate the utility of a low-cost disposable polymer microfluidic system for rapid pathogen detection for point-of-care diagnostics.
机译:本文报道了一种用于快速检测耐甲氧西林的金黄色葡萄球菌(MRSA)的聚合物微流体labon-a-slide的设计,制备和实验验证。使用磁珠将MRSA细胞捕获在裂解室中,然后进行热裂解。释放的DNA被转移到第二个反应室进行聚合酶链反应(PCR)扩增。装置中的流体控制是通过气动驱动微型泵和集成在装置上的五个微型阀来完成的。来自MRSA的mecA基因在35分钟内通过实时PCR成功扩增。使用毛细管电泳验证了正确的PCR扩增子(186 bp)的存在。结果证明了低成本的一次性聚合物微流控系统可用于快速即时病原体检测以进行即时诊断。