This paper describes the short and the long term behaviour of two different materials that have been suggested for calibration of space charge measuring systems. Two different space charge measuring systems were planned to be used for the investigation; more specifically, the laser induced pressure pulse (LIPP) system and the pulsed electroacoustic (PEA) system. The material used for the calibration test was 100 μm thick corona charged PET (Mylar) film. It was concluded that such film is unlikely to be useful for calibration purposes
机译:本文介绍了已建议用于空间电荷测量系统校准的两种不同材料的短期和长期行为。计划使用两种不同的空间电荷测量系统进行调查;更具体地说,是激光感应压力脉冲(LIPP)系统和脉冲电声(PEA)系统。用于校准测试的材料是100μm厚的带电晕的PET(聚酯薄膜)薄膜。结论是,这种膜不太可能用于校准目的