Measurement of the response of a liquid scintillation detector to monoenergetic electrons and neutrons

机译:测量液体闪烁探测器的响应   单能电子和中子

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The response of the liquid scintillator (EJ-301 equivalent to NE-213) to themonoenergetic electrons produced in Compton scattered $\gamma$-ray tagging hasbeen carried out for various radioactive $\gamma$-ray sources. The measuredelectron response is found to be linear up to $\sim$4~MeVee and the resolutionof the liquid scintillator at 1~MeVee is observed to be $\sim$~11\%. The pulseshape discrimination and pulse height response of the liquid scintillator forneutrons has been measured using $^7$Li(p,n$_1$)$^7$Be*(0.429 MeV) reaction.Non linear response to mono-energetic neutrons for the liquid scintillator isobserved at E$_n$=5.3, 9.0 and 12.7 MeV. The measured response of the liquidscintillator for electrons and neutrons have been compared with Geant4simulation.
机译:液体闪烁体(相当于NE-213的EJ-301)对康普顿散射的γ射线标记产生的单能电子的响应已针对各种放射性γ射线源进行了。发现测得的电子响应在$ \ sim $ 4〜MeVee范围内是线性的,并且在1〜MeVee处,液体闪烁体的分辨率为$ \ sim $〜11 \%。使用$ ^ 7 $ Li(p,n $ _1 $)$ ^ 7 $ Be *(0.429 MeV)反应测量液体闪烁体中子的脉冲形状判别和脉冲高度响应。液体闪烁体的观测值为E $ _n $ = 5.3、9.0和12.7 MeV。液体闪烁体对电子和中子的测量响应已与Geant4模拟进行了比较。