Automated Setup to Accurately Calibrate Electrical DC Voltage Generators

机译:自动设置以精确校准电气直流电压发生器

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

At National Institute of Metrological Research (INRIM), an automated setup tocalibrate DC Voltage generators, mainly top-level calibrators from 1 mV to 1 kVhas been developed. The heart of the setup is an INRIM-built automated fixedratios DC Voltage divider. The significant achievement of this setup is thepossibility to interconnect the divider, a DMM characterized in linearity, a DCVoltage Standard and a DC Voltage generator under calibration and automaticallyto manage the calibration process. This calibration method allows to save a lotof time, to improve the reliability and to increase the accuracy of thecalibration of generators. The relative uncertainties of the system span from0.6x10-6 to 1.2x10-4 improving the previous capabilities of the INRIMlaboratory for calibration of programmable multifunction instruments. Inaddition, this system allows to avoid the employment of several Standards (someof them still manually operating) carrying out the entire process withoutchanging the setup configuration and without the presence of operators. Theconcept of this setup can be transferred to secondary high-level electricalcalibration Laboratories that could be consider it useful for their calibrationactivities.
机译:在国家计量研究所(INRIM),已开发出一种用于校准DC电压发生器的自动设置,主要用于从1 mV到1 kV的顶级校准器。该设置的核心是INRIM内置的自动固定比率DC分压器。此设置的重大成就是可以将分压器,线性化的数字万用表,校准后的直流电压标准和直流电压发生器互连,并自动管理校准过程。这种校准方法可以节省大量时间,提高可靠性并提高发电机校准的准确性。系统的相对不确定性范围从0.6x10-6到1.2x10-4,从而提高了INRIMlaboratory用于校准多功能仪器的先前功能。此外,该系统还可以避免在执行整个过程的过程中采用多个标准(其中一些仍在手动操作),而无需更改设置配置和操作员。可以将这种设置的概念转移到二级高级电校准实验室,该实验室可能对其校准活动很有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号