Image-quality assessment for several positron emitters using the nema nu 4-2009 standards in the siemens inveon small-animal pet scanner

机译:在西门子Inveon小动物宠物扫描仪中使用nema nu 4-2009标准对多个正电子发射器的图像质量进行评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The positron emitters 18F, 68Ga, 124I, and 89Zr are all relevant in small-animal PET. Each of these radionuclides has different positron energies and ranges and a different fraction of single photons emitted. Average positron ranges larger than the intrinsic spatial resolution of the scanner (for 124I and 68Ga) will deteriorate the effective spatial resolution and activity recovery coefficient (RC) for small lesions or phantom structures. The presence of single photons (for 124I and 89Zr) could increase image noise and spillover ratios (SORs).
机译:正电子发射体18F,68Ga,124I和89Zr与小动物PET有关。这些放射性核素中的每一个都有不同的正电子能量和范围,并发射不同比例的单个光子。平均正电子范围大于扫描仪的固有空间分辨率(对于124I和68Ga)将降低小病变或体模结构的有效空间分辨率和活动恢复系数(RC)。单光子(对于124I和89Zr)的存在会增加图像噪声和溢出比(SOR)。