Apparent measurement errors in 'Development of biomechanical response corridors of the thorax to blunt ballistic impacts'

机译:“生物力学响应的发展”中的表观测量误差   胸部走廊钝性弹道影响“

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

"Development of biomechanical response corridors of the thorax to bluntballistic impacts" (Bir, C., Viano, D., King, A., 2004, Journal of Biomechanics37, 73-79.) contains apparent measurement errors. Areas under several force vs.time (Fig. 2) and force vs. deflection curves (Fig.4) differ significantly fromthe momentum and kinetic energy changes, respectively. These differencesviolate Newton's second law and the work-energy theorem.
机译:“胸腔对钝球冲击的生物力学反应通道的发展”(Bir,C.,Viano,D.,King,A.,2004,Journal of Biomechanics37,73-79。)包含明显的测量误差。在几个力与时间(图2)和力与挠度曲线(图4)下的面积分别不同于动量和动能变化。这些差异违反了牛顿第二定律和功能定理。