首页> 外文OA文献 >Determination of water in organic solvents by flow-injection analysis with Karl Fischer reagent and a biamperometric detection system
【2h】

Determination of water in organic solvents by flow-injection analysis with Karl Fischer reagent and a biamperometric detection system

机译:卡尔·费休试剂和双安培检测系统的流动注射分析法测定有机溶剂中的水

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A flow-injection system with a biamperometric flow-through detector provided with two platinum plate electrodes was tested for the determination of water with a two-component pyridine-free Karl Fischer reagent. The response was shown to be linear in the concentration range 0.03–0.11% water in methanol, ethanol or 2-propanol, with methanol as the carrier solvent. The maximum sampling frequency was about 150 samples per hr. It appeared to be possible to introduce a membrane separation step, thus allowing for the determination of water in fouled process streams. To avoid direct contact between the Karl Fischer solution and the pumping tubes, and thus extend the lifetime of the tubes, an indirect delivery system, based on replacement of the solution by pumped silicone oil, was also applied.
机译:测试了带有双安培流通检测器的流动注射系统,该检测器配有两个铂板电极,用于测定无双组分吡啶的卡尔·费休试剂中的水。在甲醇,乙醇或2-丙醇中的0.03-0.11%的水浓度范围内,以甲醇为载体溶剂,响应显示为线性。最大采样频率为每小时约150个样本。似乎可以引入膜分离步骤,从而可以确定结垢工艺物流中的水。为避免Karl Fischer溶液与泵管直接接触,从而延长管的使用寿命,还采用了间接输送系统,该系统基于泵送硅油替代溶液。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号