首页> 外文OA文献 >Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device heartmate II
【2h】

Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device heartmate II

机译:持续性左室辅助装置心伴侣II患者溶血是血栓栓塞事件和急性泵血栓形成的第一个迹象

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

textabstractBackground Despite advances in pump technology, thromboembolic events/acute pump thrombosis remain potentially life-threatening complications in patients with continuous-flow left ventricular assist devices (CF-LVAD). We sought to determine early signs of thromboembolic event/ pump thrombosis in patients with CF-LVAD, which could lead to earlier intervention. Methods We analysed all HeartMate II recipients (n = 40) in our centre between December 2006 and July 2013. Thromboembolic event/pump thrombosis was defined as a transient ischaemic attack (TIA), ischaemic cerebrovascular accident (CVA), or pump thrombosis. Results During median LVAD support of 336 days [IQR: 182–808], 8 (20%) patients developed a thromboembolic event/pump thrombosis (six TIA/CVA, two pump thromboses). At the time of the thromboembolic event/pump thrombosis, significantly higher pump power was seen compared with the no-thrombosis group (8.2 ± 3.0 vs. 6.4 ± 1.4 W, p = 0.02), as well as a trend towards a lower pulse index (4.1 ± 1.5 vs. 5.0 ± 1.0, p = 0.05) and a trend towards higher pump flow (5.7 ± 1.0 vs. 4.9 ± 1.9 L m, p = 0.06). The thrombosis group had a more than fourfold higher lactate dehydrogenase (LDH) median 1548 [IQR: 754– 2379] vs. 363 [IQR: 325–443] U/L, p = 0.0001). Bacterial (n = 4) or viral (n = 1) infection was present in 5 out of 8 patients. LDH > 735 U/L predicted thromboembolic events/ pump thrombosis with a positive predictive value of 88%. Conclusions In patients with a CF-LVAD (HeartMate II), thromboembolic events and/or pump thrombosis are associated with symptoms and signs of acute haemolysis as manifested by a high LDH, elevated pump power and decreased pulse index, especially in the context of an infection.
机译:尽管泵技术不断进步,但对于连续流左心室辅助装置(CF-LVAD)的患者,血栓栓塞事件/急性泵血栓形成仍可能危及生命。我们试图确定CF-LVAD患者的血栓栓塞事件/泵血栓形成的早期迹象,这可能导致更早的干预。方法我们分析了2006年12月至2013年7月在我们中心的所有HeartMate II接受者(n = 40)。血栓栓塞事件/泵血栓形成被定义为短暂性脑缺血发作(TIA),缺血性脑血管意外(CVA)或泵血栓形成。结果在336天的LVAD支持中值[IQR:182–808]中,有8位(20%)患者发生了血栓栓塞事件/泵血栓形成(6个TIA / CVA,2个泵血栓形成)。在血栓栓塞事件/泵血栓形成时,与无血栓形成组相比,泵浦功率明显提高(8.2±3.0 vs. 6.4±1.4 W,p = 0.02),并且脉搏指数降低(4.1±1.5 vs. 5.0±1.0,p = 0.05)和更高的泵流量趋势(5.7±1.0 vs. 4.9±1.9 L m,p = 0.06)。血栓形成组的乳酸脱氢酶(LDH)中位数为1548 [IQR:754–2379]比363 [IQR:325–443] U / L高四倍以上,p = 0.0001)。 8名患者中有5名出现细菌(n = 4)或病毒(n = 1)感染。 LDH> 735 U / L预测的血栓栓塞事件/泵血栓形成,阳性预测值为88%。结论在CF-LVAD(HeartMate II)患者中,血栓栓塞事件和/或泵血栓形成与急性溶血的症状和体征有关,表现为LDH高,泵浦功率升高和脉搏指数降低,特别是在血栓形成的情况下。感染。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号