首页> 外文OA文献 >Calibration Procedure for Fuel Flow Meters at the Nebraska Tractor Test Lab
【2h】

Calibration Procedure for Fuel Flow Meters at the Nebraska Tractor Test Lab

机译:内布拉斯加州拖拉机测试实验室燃油流量计的校准程序

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Reports in the literature indicated several factors that can influence the accuracy of Coriolis Effect mass flow meters. A Coriolis Effect mass flow meter is used to verify tractor manufacturer’s fuel consumption claims at the Nebraska Tractor Test Laboratory (NTTL). The accuracy requirement placed on the flow meter by the Organization for Economic Co-operation and Development (OECD) in the Code 2 tractor performance test procedure is not clear, but in the most conservative interpretation is ±0.5% of each flow rate measured. Results showed a dynamic weighing calibration method was not accurate enough to obtain a calibration of the flow meter to the desired accuracy level. A static weighing calibration method developed showed no significant difference between the calibration determined by the flow meter’s manufacturer with water and the calibration determined by NTTL with No. 2 diesel fuel. Static weighing calibration tests showed that for flow rates at or above 32 kg/h, the flow meter met the ±0.5% error most conservative interpretation of tolerance on flow rate from OECD Code 2.
机译:文献中的报告指出了几个可能影响科里奥利效应质量流量计精度的因素。科里奥利效应质量流量计用于在内布拉斯加州拖拉机测试实验室(NTTL)验证拖拉机制造商的油耗要求。经济合作与发展组织(OECD)在Code 2拖拉机性能测试程序中对流量计提出的精度要求尚不明确,但最保守的解释是所测得每种流量的±0.5%。结果表明,动态称重校准方法不够精确,无法将流量计校准到所需的精度水平。静态称重校准方法的开发表明,流量计制造商对水的校准与NTTL对2号柴油的校准之间没有显着差异。静态称重校准测试表明,对于等于或大于32 kg / h的流量,流量计达到了±0.5%的误差,这是OECD法规2对流量公差的最保守的解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号