A method to synchronize signals from multiple patient monitoring devices through a single input channel for inclusion in list-mode acquisitions

机译:一种通过单个输入通道同步来自多个患者监测设备的信号以包括在列表模式采集中的方法

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

PURPOSE: This technical note documents a method that the authors developed for combining a signal to synchronize a patient-monitoring device with a second physiological signal for inclusion into list-mode acquisition. Our specific application requires synchronizing an external patient motion-tracking system with a medical imaging system by multiplexing the tracking input with the ECG input. The authors believe that their methodology can be adapted for use in a variety of medical imaging modalities including single photon emission computed tomography (SPECT) and positron emission tomography (PET).METHODS: The authors insert a unique pulse sequence into a single physiological input channel. This sequence is then recorded in the list-mode acquisition along with the R-wave pulse used for ECG gating. The specific form of our pulse sequence allows for recognition of the time point being synchronized even when portions of the pulse sequence are lost due to collisions with R-wave pulses. This was achieved by altering our software used in binning the list-mode data to recognize even a portion of our pulse sequence. Limitations on heart rates at which our pulse sequence could be reliably detected were investigated by simulating the mixing of the two signals as a function of heart rate and time point during the cardiac cycle at which our pulse sequence is mixed with the cardiac signal.RESULTS: The authors have successfully achieved accurate temporal synchronization of our motion-tracking system with acquisition of SPECT projections used in 17 recent clinical research cases. In our simulation analysis the authors determined that synchronization to enable compensation for body and respiratory motion could be achieved for heart rates up to 125 beats-per-minute (bpm).CONCLUSIONS: Synchronization of list-mode acquisition with external patient monitoring devices such as those employed in motion-tracking can reliably be achieved using a simple method that can be implemented using minimal external hardware and software modification through a single input channel, while still recording cardiac gating signals.
机译:目的:该技术说明记录了作者开发的一种方法,该方法用于组合信号以将患者监护设备与第二个生理信号同步,以纳入列表模式采集。我们的特定应用需要通过将跟踪输入与ECG输入复用来使外部患者运动跟踪系统与医学成像系统同步。作者认为,他们的方法可适用于多种医学成像模式,包括单光子发射计算机断层扫描(SPECT)和正电子发射断层扫描(PET)。方法:作者将唯一的脉冲序列插入单个生理输入通道。 。然后将此序列与用于ECG门控的R波脉冲一起记录在列表模式采集中。我们的脉冲序列的特定形式允许识别同步的时间点,即使由于与R波脉冲的碰撞而丢失了部分脉冲序列。这是通过更改用于对列表模式数据进行分箱的软件来识别甚至脉冲序列的一部分来实现的。通过模拟两个信号在心动周期与心跳信号混合的心动周期中作为心率和时间点的函数的混合,研究了可以可靠地检测到我们的脉搏序列的心率限制。作者通过获取在17个最近的临床研究案例中使用的SPECT投影,成功地实现了我们的运动跟踪系统的精确时间同步。在我们的仿真分析中,作者确定可以以高达125次/分钟(bpm)的心率实现同步以补偿身体和呼吸运动。结论:列表模式采集与外部患者监测设备(例如,可以使用简单的方法可靠地实现运动跟踪中采用的那些方法,该方法可以通过单个输入通道使用最少的外部硬件和软件修改来实现,同时仍记录心脏门控信号。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号