Performance characterization of a new CZT-based preclinical SPECT system: a comparative study of different collimators

机译:基于CZT的新临床前spECT系统的性能表征:不同准直器的比较研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Triumph X-SPECT is a newly released CZT-based preclinical small-animal SPECT system with interchangeable collimators. The purpose of this work was to evaluate and systematically compare the imaging performances of three different collimators in the CZT-based preclinical small-animal system: a single-pinhole collimator (SPH), a multi-pinhole collimator (MPH) and a parallel-hole collimator. We measured the spatial resolutions and sensitivities of the three collimators with Tc-99m sources, considering three distinct energy window widths (5, 10, and 20%), and used the NEMA NU4-2008 Image Quality phantom to test the imaging performance of the three collimators in terms of uniformity and spill-over ratio (SOR) for each energy window. With a 10% energy window width at a radius of rotation (ROR) of 30 mm, the system resolution of the SPH, MPH and parallel-hole collimators was 0.715, 0.855 and 3.270 mm FWHM, respectively. For the same energy window, the sensitivity of the system with SPH, MPH and parallel-hole collimators was 32.860, 152.514 and 49.205 counts/sec/MBq at a 100 mm source-to-detector distance and 6.790, 33.376 and 49.038 counts/sec/MBq at a 130 mm source-to-detector distance, respectively. The image noise and SORair for the three collimators were 20.137, 12.278 and 11.232 (%STDunif) and 0.106, 0.140 and 0.161, respectively. Overall, the results show that the SPH had better spatial resolution than the other collimators. The MPH had the highest sensitivity at 100 mm source-to-collimator distance, and the parallel-hole collimator had the highest sensitivity at 130 mm-source-to-detector distance. Therefore, the proper collimator for Triumph X-SPECT system must be determined by the task. These results provide valuable reference data and insight into the imaging performance of various collimators in CZT-based preclinical small-animal SPECT.
机译:胜利X-SPECT是新发布的基于CZT的临床前小动物SPECT系统,具有可互换的准直器。这项工作的目的是评估和系统比较基于CZT的临床前小动物系统中三种不同准直仪的成像性能:单针孔准直仪(SPH),多针孔准直仪(MPH)和平行针准直仪。孔准直器。考虑到三个不同的能量窗口宽度(5%,10%和20%),我们使用Tc-99m光源测量了三个准直仪的空间分辨率和灵敏度,并使用NEMA NU4-2008图像质量幻像来测试图像的成像性能。在每个能量窗口的均匀度和溢出率(SOR)方面三个准直器。当旋转半径(ROR)为30 mm时,能量窗口宽度为10%时,SPH,MPH和平行孔准直仪的系统分辨率分别为FWHM 0.715、0.855和3.270 mm。对于相同的能量窗口,使用SPH,MPH和平行孔准直仪的系统在100 mm的源到检测器距离和6.790、33.376和49.038个计数/秒下的灵敏度分别为32.860、152.514和49.205个计数/秒/ MBq。 / MBq分别以130 mm的源到检测器距离。三个准直器的图像噪声和SORair分别为20.137、12.278和11.232(%STDunif)和0.106、0.140和0.161。总体而言,结果表明SPH的空间分辨率高于其他准直仪。 MPH在源到准直仪距离为100 mm时灵敏度最高,而平行孔准直仪在源到检波仪距离为130 mm时灵敏度最高。因此,必须根据任务确定适合Triumph X-SPECT系统的准直仪。这些结果为基于CZT的临床前小动物SPECT中各种准直仪的成像性能提供了有价值的参考数据。

著录项

  • 作者

    ???;

  • 作者单位
  • 年度 2015
  • 总页数
  • 原文格式 PDF
  • 正文语种 en
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号