首页> 外文OA文献 >Simultaneous Determination of Carboxylic Acids and Inorganic Anions by Photometric Ion Chromatography Application to Liquid Foods
【2h】

Simultaneous Determination of Carboxylic Acids and Inorganic Anions by Photometric Ion Chromatography Application to Liquid Foods

机译:光度离子色谱法同时测定液体食品中羧酸和无机阴离子的含量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Simultaneous determination of carboxylic acids and inorganic anions was investigated by photometric ion chromatography with phthalate solution as an eluent. The retention time and the peak area of carboxylic acids changed remarkably with eluent pH. The optimum conditions at present were as follows : analytical column, OYOBUNKO ASA-4000 (4.6 mm i.d.×25 cm) ; eluent, 5×10-4 M disodium phthalate ; flow rate, 2 ml/min ; column temperature, 40°C ; detection wavelength, 240 nm ; injection volume, 100 μl. The detection limits for several carboxylic acids as 3 times noise were 0.05-0.07 mg/l in sample solution. Their calibration curves showed straight lines by the peak area method. The present method has an advantage to determine simultaneously not only carboxylic acids but also inorganic anions. Under the condition described above, both carboxylic acids and inorganic anions in several liquid foods were easily determined by direct injection of the sample prepared with only dilution.
机译:以邻苯二甲酸盐溶液为洗脱液,通过光度离子色谱法同时测定了羧酸和无机阴离子。羧酸的保留时间和峰面积随洗脱液pH的变化而显着变化。目前的最佳条件如下:OYOBUNKO ASA-4000分析柱(内径4.6 mm×25 cm);洗脱液,5×10-4 M邻苯二甲酸二钠;流速2 ml / min柱温40°C;检测波长240 nm;进样量100μl样品溶液中几种羧酸的检测限是噪声的3倍,为0.05-0.07 mg / l。其校正曲线通过峰面积法显示为直线。本方法的优点在于不仅可以同时测定羧酸,而且还可以测定无机阴离子。在上述条件下,通过直接注射仅稀释制备的样品,即可轻松测定几种液体食品中的羧酸和无机阴离子。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号