The Finapres 2300e finger cuff. The influence of cuff application on the accuracy of blood pressure measurement

机译:Finapres 2300e手指套。袖带应用对血压测量精度的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Blood pressure measurements associated with malapplication of a finger cuff were compared with contemporaneous intra-arterial pressure data in seven volunteers to determine the influence of cuff application on the accuracy of the Finapres 2300e. Systolic readings in all cuff positions differed from arterial line data by more than the recommended standard and tight and loose cuff applications under and over-read respectively, in all subjects. The results show the Finapres to be sensitive to small degrees of finger cuff malapplication which contribute to the bias on direct arterial comparison and limit the reliability of the instrument in clinical practice.
机译:在七个志愿者中,将与手指套箍使用不当相关的血压测量结果与同期动脉内压力数据进行了比较,以确定套箍套施加对Finapres 2300e准确性的影响。在所有受试者中,所有袖带位置的收缩压读数与动脉线数据的差异均超过建议的标准,在低于和超过读数的情况下,袖带的紧绷和松弛情况均超过推荐值。结果表明,Finapres对手指套箍的小程度应用不灵敏,这会导致直接动脉比较出现偏差,并限制了该器械在临床实践中的可靠性。