首页> 外文OA文献 >Tidal volume monitoring by a set of tetrapolar impedance measurements selected from the 16-electrodes arrangement used inelectrical impedance tomography (EIT) technique. Calibration equations in a group of healthy males
【2h】

Tidal volume monitoring by a set of tetrapolar impedance measurements selected from the 16-electrodes arrangement used inelectrical impedance tomography (EIT) technique. Calibration equations in a group of healthy males

机译:通过选自使用电阻抗断层摄影(EIT)技术的16电极布置的一组四极阻抗测量来监测潮气量。一组健康男性的校准方程

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Previously, our research group obtained a set of calibration equations for being used by an electricalimpedance tomograph (EIT). All equations transform the impedance changes into a measurable volumesignal in a group of healthy males. The performance of EIT, using these equations, was acceptable forrespiration monitoring. The EIT impedance changes were obtained from a set of 16-electrodes placedaround the thoracic box at level of the sixth intercostal space. Each cycle of impedance measurementsis ordered in a matrix (IEITM). Each IEITM’s element depicts a configuration of 4-electrodes. Now, ourmain challenge is to replace the EIT’s 16-electrodes by a 4-electrodes configuration. We analyzed theimpedance changes obtained from each element and the volume determinations obtained by a pneu-motachometer (gold standard) in order to determine the optimal 4-electrodes configuration. For eachselected configuration a set of 20 calibration equations were obtained. The best results were obtained byusing two 4-electrodes configurations. Both consisting in two adjacent electrodes for current injectionand two electrodes for voltage detection.The mean R2of the 20 equations determined by the 16-electrodes information, and for the best two 4-electrode configurations were 0.943 ± 0.010, 0.848 ± 0.062 and 0.690 ± 0.122, respectively. The error (%)of volume determinations obtained by the use of 16-electrodes, and by the use of the two best 4-electrodeconfiguration with regarding to the real volume obtained by pneumotachometer were of 15 ± 6%, 16 ± 4%and 43 ± 41%, respectively. The volume differences between the one obtained by the use of 16-electrodesand that for the best 4 electrode configuration were not significant. We conclude that although thedeterminations of impedance obtained by the best 4-electrode configuration has a lower sensitivity thanthose obtained by the 16-electrodes, it is possible to measure the respiratory pattern in healthy males
机译:以前,我们的研究小组获得了一组用于电抗断层扫描仪(EIT)的校准方程式。所有方程式将一组健康男性的阻抗变化转换为可测量的体积信号。使用这些公式,EIT的性能对于呼吸监测是可以接受的。 EIT阻抗变化是从一组位于第六肋间水平的胸腔周围的16电极获得的。阻抗测量的每个周期都按矩阵(IEITM)排序。每个IEITM元素都描绘了一个4电极的配置。现在,我们的主要挑战是将EIT的16电极替换为4电极配置。为了确定最佳的4电极配置,我们分析了从每个元素获得的阻抗变化以及通过气动测速仪(金标准)获得的体积确定值。对于每个选定的配置,获得了20个校准方程式的集合。通过使用两个4电极配置可获得最佳结果。两者都由两个相邻的用于注入电流的电极和两个用于检测电压的电极组成。由16电极信息确定的20个方程的平均值R2以及最佳的两个4电极配置分别为0.943±0.010、0.848±0.062和0.690±0.122 , 分别。通过使用16电极以及相对于通过气压计获得的实际体积使用两种最佳的4电极配置获得的体积测定的误差(%)为15±6%,16±4%和43分别为±41%。通过使用16电极获得的体积与最佳4电极配置获得的体积差异不明显。我们得出结论,尽管通过最佳4电极配置获得的阻抗测定比通过16电极获得的阻抗灵敏度低,但是可以测量健康男性的呼吸模式

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号