A new detection system for the measurement of beta-delayed neutron emission probabilities has been characterized using fission products with well known ß-delayed neutron emission properties. The setup consists of BELEN-20, a 4p neutron counter with twenty 3He proportional tubes arranged inside a large polyethylene neutron moderator, a thin Si detector for ß counting and a self-triggering digital data acquisition system. The use of delayed-neutron precursors with different neutron emission windows allowed the study of the effect of energy dependency on neutron, ß and ß-neutron rates. The observed effect is well reproduced by Monte Carlo simulations. The impact of this dependency on the accuracy of neutron emission probabilities is discussed. A new accurate value of the neutron emission probability for the important delayed-neutron precursor 137I was obtained, Pn=7.76(14)%.
机译:使用具有众所周知的ß延迟中子发射特性的裂变产物,表征了一种用于测量β延迟中子发射概率的新检测系统。该装置包括BELEN-20,一个4p中子计数器和一个布置在大型聚乙烯中子减速器内的20He 3He比例管,一个用于ß计数的薄硅探测器以及一个自触发数字数据采集系统。使用具有不同中子发射窗口的延迟中子前体可以研究能量依赖性对中子,ß和ß中子速率的影响。蒙特卡罗模拟很好地再现了观察到的效果。讨论了这种依赖性对中子发射概率准确性的影响。获得了重要的延迟中子先驱体137I的中子发射概率的新精确值,Pn = 7.76(14)%。