Implantable medical devices, such as stents, have to be inspected 100% so no defective ones are implanted into a human body. In this paper, a novel optical stent inspection system is presented. By the combination of a high numerical aperture microscope, a triple illumination system, a rotational stage, and a CMOS camera, unrolled sections of the outer and inner surfaces of the stent are obtained with high resolution at high speed with a line-scan approach. In this paper, a comparison between the conventional microscope image formation and this new approach is shown. A calibration process and the investigation of the error sources that lead to inaccuracies of the critical dimension measurements are presented.
机译:诸如支架之类的可植入医疗设备必须进行100%的检查,因此不能将有缺陷的医疗设备植入人体。在本文中,提出了一种新颖的光学支架检查系统。通过将高数值孔径显微镜,三重照明系统,旋转平台和CMOS相机相结合,可以通过线扫描方法以高分辨率获得支架外表面和内表面的展开部分。在本文中,显示了常规显微镜成像与该新方法之间的比较。给出了导致临界尺寸测量不准确的校准过程和误差源调查。