Model based iterative reconstruction IMR gives possibility to evaluate thinner slice thicknesses than conventional iterative reconstruction iDose(4) - a phantom study

机译:基于模型的迭代重建ImR提供了评估比传统迭代重建更薄的切片厚度的可能性iDose(4) - 一个幻像研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Computed tomography (CT) is one of the most important modalities in a radiological department, which produces images with high diagnostic confidence, but in some cases contributes to a high radiation dose to the patient. The radiation dose can be reduced by the use of advanced image reconstruction algorithms. This study was done on a Philips Brilliance iCT with iterative reconstruction iDose(4) and model-based iterative reconstruction IMR. The purpose was to investigate the effect on the image quality with thin slice images reconstructed with IMR, compared to standard slice thickness reconstructed with iDose(4). Objective measurements of noise and contrast-to-noise ratio were performed using an image quality phantom, an anthropomorphic phantom and clinical cases. Subjective evaluations of low-contrast resolution were performed by observers using an image quality phantom. IMR gives strong noise reduction and enhanced low-contrast and thereby enable selection of thinner slice thickness. Objective evaluation of image noise shows that thin slices reconstructed with IMR provides lower noise than thicker slice images reconstructed with iDose(4). With IMR the slice thickness is of less importance for the noise. With thinner slices the partial volume artefacts becomes less pronounced. In conclusion, we have shown that IMR enables reduction of the slice thickness and at the same time maintain or even reduce the noise level compared to iDose(4) reconstruction with standard slice thickness. This will subsequently result in an improvement of image quality for images reconstructed with IMR.
机译:计算机断层扫描(CT)是放射科最重要的方式之一,它可以产生具有高诊断可信度的图像,但在某些情况下会给患者带来高辐射剂量。辐射剂量可以通过使用高级图像重建算法来降低。这项研究是在具有迭代重建iDose(4)和基于模型的迭代重建IMR的Philips Brilliance iCT上完成的。目的是研究与使用iDose(4)重建的标准切片厚度相比,使用IMR重建的薄片图像对图像质量的影响。使用图像质量体模,拟人化体模和临床病例对噪声和对比度-噪声比进行客观测量。观察者使用图像质量体模对低对比度分辨率进行主观评估。 IMR可以大大降低噪声并增强低对比度,从而可以选择更薄的切片厚度。图像噪声的客观评估表明,与使用iDose(4)重建的较厚切片图像相比,使用IMR重建的较薄切片具有较低的噪声。使用IMR,切片厚度对于噪声的重要性就不那么重要了。切片越薄,部分体积的伪影就越不明显。总之,我们已经证明,与使用标准切片厚度的iDose(4)重建相比,IMR可以减少切片厚度,并同时保持甚至降低噪声水平。随后,这将导致使用IMR重建的图像的图像质量得到改善。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号