Intraobserver and interobserver variability of ovarian volume, gray-scale and color flow indices obtained using transvaginal three-dimensional power Doppler ultrasonography

机译:使用经阴道三维能量多普勒超声检查获得的卵巢体积,灰度和彩色血流指数的观察者和观察者间变异性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective To assess intraobserver and interobserver variability in ovarian volume and gray-scale and color flow index measurements using transvaginal, three-dimensional, power Doppler ultrasonography. Methods Eleven women (22 ovaries) were examined on day 8 of controlled ovarian hyperstimulation therapy, which was part of their in vitro fertilization treatment protocol. The patients were examined twice by the first observer and once by the second observer. The acquired volume datasets were analyzed using the VOCAL(TM) imaging program, enabling the assessment of ovarian volume, vascularization index (VI), flow index (FI), vascularization flow index (VFI) and mean grayness (MG). For these parameters the intraclass (intra-CC) and interclass (inter-CC) correlation coefficients, within-observer and between-observers repeatability coefficient (r) and limits of agreement were calculated. Results Both intraobserver and interobserver repeatability of ovarian volume measurements were considered very good with an intra-CC value of 1.00 and inter-CC value of 0.99, respectively. Also VI, FI, VFI and MG measurements were repeatable by a single observer, the intra-CC ranging from 0.82 to 0.91. The interobserver reproducibility was also good for VI, VFI and MG measurements (inter-CC values 0.73, 0.70 and 0.81, respectively), but for FI measurements the reproducibility was poor (inter-CC = 0.29, r = 7.87). Conclusions In general, the intraobserver reproducibility was better than interobserver reproducibility for all parameters. The volume assessments were reproducible both by one observer and by two separate observers. The intraobserver and interobserver variabilities were acceptable for VI, VFI and MG, whereas for FI the interobserver reproducibility was poor. Our results suggest that measurement of gray-scale and color Doppler flow indices is reproducible thus allowing them to be used in clinical practice and research.
机译:目的评估经阴道,三维,功率多普勒超声检查对卵巢体积的观察者内和观察者间变异性以及灰度和色流指数的测量。方法在第8天进行了控制性卵巢过度刺激治疗的11名妇女(22个卵巢),这是她们体外受精治疗方案的一部分。第一位观察者对患者进行了两次检查,第二位观察者对患者进行了一次检查。使用VOCAL™成像程序分析采集的体积数据集,从而能够评估卵巢体积,血管化指数(VI),血流指数(FI),血管化血流指数(VFI)和平均灰度(MG)。对于这些参数,计算了类内(CC内)和类间(CC间)的相关系数,观察者内部和观察者之间的重复性系数(r)和一致性极限。结果卵巢体积测量的观察者内和观察者间重复性都被认为非常好,CC内值为1.00,CC间值为0.99。 VI,FI,VFI和MG的测量也可以由单个观察者重复进行,CC内的范围为0.82至0.91。观察者之间的可重复性对于VI,VFI和MG测量也很好(CC间值分别为0.73、0.70和0.81),但对于FI测量,可重复性很差(CC间= 0.29,r = 7.87)。结论通常,在所有参数下,观察者内部的可复制性均优于观察者之间的可复制性。体积评估可由一名观察员和两名独立观察员进行。 VI,VFI和MG的观察者间和观察者间差异是可以接受的,而FI的观察者间再现性较差。我们的结果表明,灰度和彩色多普勒血流指数的测量是可重复的,因此可以将它们用于临床实践和研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号