A two-channel, thermal dissociation cavity-ringdown spectrometer for the detection of ambient NO2, RO2NO2 and RONO2

机译:用于检测环境NO 2 ,RO 2 NO 2 和RONO 2的双通道热解离腔衰荡光谱仪

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We describe a Thermal Dissociation Cavity-Ring-Down Spectrometer (TD-CRDS) for measurement of ambient NO2, total peroxy nitrates (ŒPNs) and total alkyl nitrates (ŒANs). The spectrometer has two separate cavities operating at 405.2 and 408.5 nm, one cavity (reference) samples NO2 continuously from an inlet at ambient 5 temperature, the other samples sequentially from an inlet at 473 K in which PNs areconverted to NO2 or from an inlet at 723 K in which both PNs and ANs are converted to NO2, dierence signals being used to derive mixing ratios of ŒPNs andŒANs. We describe an extensive set of laboratory experiments and numerical simulations to characterise the fate of organic radicals in the hot inlets and cavity and derive correction 10 factors to account for the bias resulting from interaction of peroxy radicals with ambient NO and NO2 . Finally, we present the first measurements and comparison with other instruments during a field campaign, outline the limitations of the present instrument and provide an outlook for future improvements.
机译:我们描述了一种热解离腔衰荡光谱仪(TD-CRDS),用于测量环境NO2,总过氧硝酸盐(ŒPNs)和总烷基硝酸盐(ŒANs)。光谱仪有两个分别在405.2和408.5 nm处工作的腔体,一个腔体(参考)在环境温度为5的温度下从入口连续采样NO2,其他腔室从473 K处的进气口依次采样,其中PNs转换为NO2或从进气口开始。 723 K,其中PN和AN均转化为NO2,差信号用于推导PN和AN的混合比。我们描述了一系列广泛的实验和数值模拟,以表征热入口和空腔中有机自由基的命运,并得出校正10个因素以解释过氧自由基与环境NO和NO2相互作用所引起的偏差。最后,在野战期间,我们将进行首次测量并与其他仪器进行比较,概述本仪器的局限性,并对未来的改进提供展望。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号