Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm.

机译:使用散射校正算法改善锥束CT数一致性的可行性。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The study was to explore the feasibility of improving cone-beam CT (CBCT) number (corresponding to the Hounsfield units in computed tomography) consistency using a scatter-correction algorithm, with the aim of using CBCT images for treatment planning with density correction. A scatter correction algorithm was applied to a Varian OBI CBCT and an Elekta XVI CBCT, and was evaluated for improving CBCT number consistency. CBCT numbers of phantom materials were compared between images with and without bolus, which introduced additional scatter, and with and without scatter correction processing. It was observed that CBCT numbers were different in the images with and without bolus in both CBCT studies, and the differences were reduced remarkably after scatter-correction processing. Results showed that CBCT number consistency was significantly improved by use of the scatter-correction algorithm.
机译:该研究旨在探讨使用散点校正算法提高锥束CT(CBCT)数量(对应于计算机断层扫描中的Hounsfield单位)一致性的可行性,目的是使用CBCT图像进行密度校正治疗计划。将散点校正算法应用于Varian OBI CBCT和Elekta XVI CBCT,并进行了评估,以改善CBCT编号一致性。在带和不带推注的图像之间比较了幻像材料的CBCT数量,这会引入额外的散射,以及带和不带散射校正处理。在两个CBCT研究中,观察到在有推注和不推注的图像中,CBCT数均不同,并且在进行散射校正处理后,差异显着减小。结果表明,通过使用散射校正算法,CBCT编号一致性得到了显着改善。