Analysis of uncertainties in dose distribution due to geometrical deviation of target from isocenter in serial tomotherapy

机译:连续断层治疗中靶标与等中心点几何偏差引起的剂量分布不确定性分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The tomotherapy system in radiation therapy department of “V. Fazzi” hospital in Lecce (Italy) is a serial device. It means that the coverage of the target is obtained by moving the couch in and out the gantry position. It has been shown that the choice of target positioning respect to the isocenter reflects on dose distribution in terms of heterogeneities. The entity of the heterogeneities can reach 10% in abutment regions nearby the isocenter. Moreover, by using gantry rotations less than 300 degrees, hot and cold spots can be seen above and over the isocenter. Aim of the present work is to characterize the performance of the serial tomotherapy device in order to avoid, as more as possible, heterogeneities in the abutment regions.
机译:五,放射治疗科的断层扫描系统位于意大利莱切的Fazzi医院是一台串行设备。这意味着可以通过将卧榻移入和移出龙门架位置来获得目标的覆盖范围。已经表明,相对于等中心点的目标定位的选择反映了剂量分布的异质性。在等中心点附近的基台区域中,异质性的实体可以达到10%。此外,通过使用小于300度的龙门旋转,可以在等角点上方和上方看到热点和冷点。本发明的目的是表征串联断层治疗装置的性能,以便尽可能避免在基台区域中的异质性。