A novel breast ultrasound system for providing coronal images : system development and feasibility study

机译:提供冠状图像的新型乳房超声系统:系统开发和可行性研究

摘要

Breast ultrasound images along coronal plane contain important diagnosis information. However, conventional clinical 2D ultrasound cannot provide such images. In order to solve this problem, we developed a novel ultrasound system aimed at providing breast coronal images. In this system, a spatial sensor was fixed on an ultrasound probe to obtain the image spatial data. A narrow-band rendering method was used to form coronal images based on B-mode images and their corresponding spatial data. Software was developed for data acquisition, processing, rendering and visualization. In phantom experiments, 20 inclusions with different size (5-20 mm) were measured using this new system. The results obtained by the new method well correlated with those measured by a micrometer (y = 1.0147x, R-2 = 0.9927). The phantom tests also showed that this system had excellent intra-and inter-operator repeatability (ICC > 0.995). Three subjects with breast lesions were scanned in vivo using this new system and a commercially available three-dimensional (3D) probe. The average scanning times for the two systems were 64 s and 74 s, respectively. The results revealed that this new method required shorter scanning time. The tumor sizes measured on the coronal plane provided by the new method were smaller by 5.6-11.9% in comparison with the results of the 3D probe. The phantom tests and preliminary subject tests indicated the feasibility of this system for clinical applications by providing additional information for clinical breast ultrasound diagnosis.
机译:沿冠状平面的乳房超声图像包含重要的诊断信息。但是,常规的临床2D超声无法提供此类图像。为了解决这个问题,我们开发了一种新颖的超声系统,旨在提供乳腺冠状图像。在该系统中,将空间传感器固定在超声探头上以获得图像空间数据。窄带渲染方法用于基于B模式图像及其相应的空间数据形成冠状图像。开发了用于数据采集,处理,渲染和可视化的软件。在幻像实验中,使用此新系统测量了20个大小不同(5-20​​毫米)的夹杂物。通过新方法获得的结果与用千分尺测量的结果高度相关(y = 1.0147x,R-2 = 0.9927)。幻像测试还表明,该系统具有出色的操作员内部和操作员之间的可重复性(ICC> 0.995)。使用此新系统和市售的三维(3D)探针在体内扫描了三个患有乳腺病变的受试者。两个系统的平均扫描时间分别为64 s和74 s。结果表明,这种新方法需要更短的扫描时间。与3D探针的结果相比,新方法在冠状面上测量的肿瘤尺寸小了5.6-11.9%。体模测试和初步主题测试通过为临床乳房超声诊断提供其他信息,表明了该系统在临床应用中的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号