Structural analysis and optimization of the support device used for a proximal fracture of the femur

机译:用于股骨近端骨折的支撑装置的结构分析和优化

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The support system for a proximal fracture of the femur is studied by means of the finite element method. The support system of the Gamma III by Stryker is used for this study due to the versatility and simplicity of the system. The variance of each system was modeled to validate the results. Parametric studies between the models were conducted and design improvements were then tested against the existing parameters. Given the material properties for the Ti-6 AL-4V support system and bone, the models were considered under two different load conditions. From these results, a recommendation can be made to the manufacturing and medical community as to the effective use of the Gamma III nail implant system for a proximal fracture in the femur.
机译:利用有限元方法研究了股骨近端骨折的支撑系统。由于系统的多功能性和简单性,Stryker的Gamma III支撑系统被用于本研究。对每个系统的方差建模以验证结果。在模型之间进行了参数研究,然后针对现有参数测试了设计改进。考虑到Ti-6 AL-4V支撑系统和骨骼的材料特性,在两种不同的载荷条件下考虑了模型。从这些结果,可以建议制造商和医学界有效使用Gamma III钉植入系统治疗股骨近端骨折。