首页> 外文OA文献 >Accuracy and repeatability of a new portable ultrasound pachymeter
【2h】

Accuracy and repeatability of a new portable ultrasound pachymeter

机译:新型便携式超声波测厚仪的准确性和可重复性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

To assess the accuracy and repeatability of central corneal thickness (CCT) measurements taken with a new portable ultrasound (US) pachymeter.Methods: Central thickness measurements were taken with a portable pachymeter (SP-100 Handy;Tomey, Nagoya, Japan) and a conventional US pachymeter (Nidek UP-1000; Nidek Technologies, Gamagori, Japan) from 57 right corneas of 57 young adults (19 males, 38 females) aged 18–44 years (mean ± S.D., 22.95 ± 3.92). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between the pachymeters. The three readings taken with the portable pachymeter were compared against each other in order to evaluate intra-session repeatability and bias of each individual measurement with respect to the mean of three.Results: Mean values of CCT were 537 ± 35 micron for conventional and 534 ± 35 micron for the new portable pachymeter. A high agreement was found between the two instruments mean difference = 2.58 micron; 95% CI 1.41–3.75 micron) with only two eyes presenting differences larger than ± 8.6 micron which represents 95% CI in the Bland-Altman plots which represents 1.6% of the mean CCT. The first reading taken showed the highest agreement with the mean value for the portable pachymeter.Conclusions: The instrument tested in this study is able to take reliable measurements of corneal thickness even if a single reading is considered. Intra-session repeatability was very high, as was also the agreement between the average of three readings taken with the two US pachymeters.
机译:为了评估使用新型便携式超声波(US)测厚仪进行的中央角膜厚度(CCT)测量的准确性和可重复性。方法:使用便携式测厚仪(SP-100 Handy;日本名古屋市通美市(Toomey),日本名古屋市)进行中心厚度测量。常规的美国测速仪(Nidek UP-1000; Nidek Technologies,日本蒲郡),年龄为18-44岁的57位年轻人(19位男性,38位女性)的57个右角膜(平均值±SD,22.95±3.92)。获得了三个重复的测量值,然后进行比较以获得每个仪器的重复性以及测厚仪之间的一致性。将便携式测厚仪获取的三个读数进行相互比较,以评估会话间的可重复性和每个单独测量相对于三个平均值的偏差。结果:CCT的平均值为常规的537±35微米和534。新的便携式测厚仪为±35微米。两种仪器之间的高度一致性为2.58微米; 95%CI为1.41–3.75微米),只有两只眼睛的差异大于±8.6微米,在Bland-Altman图中代表95%CI,占平均CCT的1.6%。首次读数显示出与便携式测厚仪的平均值的最高一致性。结论:即使考虑单个读数,本研究中测试的仪器也能够进行可靠的角膜厚度测量。会话中的可重复性非常高,两个美国测厚仪的三个读数的平均值之间的一致性也很高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号