Wearable biopotential measurement using the TI ADS1198 analog front-end and textile electrodes : signal conditioning and signal quality assessment

机译:使用TI ADS1198模拟前端和纺织电极进行可穿戴生物电势测量:信号调节和信号质量评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The development of mobile systems for monitoring bioelectric signals outside a hospital environment involves many challenges that do not arise when it is in a controlled environment, like a hospital. The dimensions of these systems are an important factor to consider in order to facilitate their use without interfering with the daily activities of individuals.The purpose of this work is the implementation of a single-supply battery-powered, low power ECG/EMG signal monitoring system based on the ADS1198 Analog Front-End from Texas Instruments. The system was designed to acquire ECG signals from three electrodes using the integrated Right-Leg-Drive (RLD) circuit from the ADS1198. The developed analog front-end was connected for testing purposes through the SPI interface to a NI-USB 8451 board and signals were acquired using LabVIEW. The circuit was tested in several situations and proved to provide high quality signals using textile integrated electrodes and conventional disposable gel electrodes.
机译:用于在医院环境之外监视生物电信号的移动系统的开发涉及许多挑战,而这些挑战在医院等可控环境中不会出现。这些系统的尺寸是在不干扰个人日常活动的前提下要考虑使用的重要因素。这项工作的目的是实现单电源电池供电的低功率ECG / EMG信号监控系统基于德州仪器(TI)的ADS1198模拟前端。该系统设计为使用ADS1198的集成右腿驱动(RLD)电路从三个电极获取ECG信号。为了测试目的,将开发的模拟前端通过SPI接口连接到NI-USB 8451板上,并使用LabVIEW采集信号。该电路已在几种情况下进行了测试,并证明使用纺织品集成电极和常规一次性凝胶电极可提供高质量信号。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号