Pengaruh Variasi Air Gap terhadap Dosis Serap Penyinaran Berkas Elektron pada Pesawat Linac Siemens / Primus M Class 5633

机译:气隙变化对西门子Linac / Primus M 5633级电子束辐照剂量的影响

摘要

The effect of the variation of the air gap absorbed dose in electron beam irradiation has conducted. In the particular case of electron beam irradiation often can not use the SSD 100 cm due to the influence of the use of the applicator so that the patient\u27s body should be granted an extension of the SSD or air gap at the irradiation. In this study the calculation of the effective SSD, absorbed dose, the air gap correction factor and verification of absorbed dose was performed using the inverse square law and the correction of the output electrons. The method utilized an inverse square law load value of each electron beam irradiation to determine the effective SSD, while the determination of absorbed dose and the air gap correction factor correction using electron output. The results of the calculation states that the effective SSD of 82,2 cm – 109 cm wide field size and is proportional to the electron energy. Absorbed dose and the air gap correction factor proportional to the field size and inversely proportional to the variation of the air gap while the value of the absorbed dose itself is varied between 64,80 cGy to 108,0 cGy as well as the value of the correction factor of the air gap is between 0,869 – 1,000. Correction dose verification of absorbed dose value is still below the value specified tolerances AAPM TG40 is equal to 99,10% - 103,0%. Variations of growing air gap causes the smaller absorbed dose.
机译:已经进行了电子束照射中气隙吸收剂量的变化的影响。在特定的电子束照射情况下,由于使用施加器的影响,SSD通常无法使用100厘米,因此在照射时应允许患者身体扩展SSD或气隙。在这项研究中,使用平方反比定律和输出电子的校正来计算有效SSD,吸收剂量,气隙校正因子并验证吸收剂量。该方法利用每次电子束辐照的平方反比负荷值确定有效的SSD,同时使用电子输出确定吸收剂量并校正气隙校正因子。计算结果表明,有效SSD的场尺寸为82.2 cm-109 cm,与电子能量成正比。吸收剂量和气隙校正因子与场大小成正比,与气隙的变化成反比,而吸收剂量本身的值在64,80 cGy至108,0 cGy之间以及气隙的校正系数在0,869 – 1,000之间。吸收剂量值的校正剂量验证仍低于规定的值,AAPM TG40的公差等于99.10%-103.0%。气隙的变化会导致较小的吸收剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号