Background: The accuracy and precision of any instrument should not be taken for granted. While there is an international standard for checking focimeters, there is no report of any study on their performance.Method: A sample set of 51 focimeters (11 brands), were used to measure the sphericalpower of a set of lenses and the prismatic power of two lenses complying with ISO9342-1:2005 and other calibrated prismatic lenses and the spherical power of some greyfilters.Results: The mean measured spherical power corresponded very closely with the calibrated values; however, the spread of results was substantial and 10 focimeters did not comply with ISO 8598:1996. The measurement of prism was much more accurate and precise and all the focimeters complied easily. With the grey filters, about one-third of the foci meters either showed erratic reading or an error with the equivalent of category 4 sunglasses. On the other hand, nine focimeters had stable and accurate reading on a filter with a luminous transmittance of 0.5 per cent.Conclusions: These results confirm that, in common with all other measurement instruments, there is a need to ensure that a focimeter is reading accurately and precisely over the range of refractive powers and luminous transmittances. The accurate and precise performance of an automated focimeter over its working life cannot be assumed. Checking before purchase with a set of calibrated lenses and some dark sunglass tints will indicate the suitability of a focimeter. Routine checking with the calibrated lenses will inform the users if a focimeter continues to indicate accurately.
机译:背景:任何仪器的准确性和精确性均不应视为理所当然。尽管有国际上的焦度计检查标准,但尚未对其性能进行任何研究的报道方法:方法:使用51个焦度计(11个品牌)的样本集来测量一组镜片的球面屈光力和屈光度的棱镜度。结果:测得的平均球面度与校准值非常接近;两片符合ISO9342-1:2005的镜片和其他校准的棱镜镜片,以及一些灰度滤镜的球面度。然而,结果的传播是巨大的,并且10英尺/分米不符合ISO 8598:1996。棱镜的测量更加准确和精确,并且所有的焦度计都易于安装。使用灰色滤光片时,约有三分之一的焦距表读数不稳定或与4类太阳镜相当。另一方面,九个光度计在透射率为0.5%的滤光片上具有稳定且准确的读数。结论:这些结果证实,与所有其他测量仪器一样,需要确保光度计正在读取在屈光力和透光率的范围内准确而精确地测量。不能假定自动聚焦计在其整个使用寿命中都具有准确和精确的性能。在购买之前,请先使用一组校准的镜片和一些深色的太阳镜色调进行检查,这将表明测光仪的适用性。如果用聚焦计继续正确显示,则用校准过的镜片进行例行检查将通知用户。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号