首页> 外文OA文献 >Rapid pre-tension loss in the Ilizarov external fixator: an in vitro study.
【2h】

Rapid pre-tension loss in the Ilizarov external fixator: an in vitro study.

机译:Ilizarov外固定器的快速预张力丧失:一项体外研究。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BACKGROUND AND PURPOSE: Wire pre-tension in the Ilizarov frame is considered to be important in order to reduce movements that can impair fracture healing. Wires will eventually lose part of their pre-tension, however. In order to gain more insight into the need for wire pre-tension, we investigated: (1) the amount of pre-ten-sion loss, (2) the relationship between pre-tension loss and bolt-tightening torque, (3) the cause of a possible loss of pre-tension, and (4) the effect of pre-tension on cyclic micromotions and total displacement of the bone segments. METHODS: 3 Ilizarov configurations, with various wire pre-tensions and bolt-tightening torques, were tested 3 times with an MTS machine. For each wire, slippage and tensions were measured for the duration of the whole experiment. RESULTS: A loss of wire pre-tension (up to 75%) due to slippage was found. Higher bolt-tightening torques significantly reduced the amount of pre-tension loss. Furthermore, a higher wire pre-tension reduced the maximal axial displacement of the bone fragment. There was, however, least cyclic axial micromotion when no wire pre-tension was present -- probably due to the lack of wire recoil. INTERPRETATION: Wires in an experimental Ilizarov external fixator lose pre-tension within a limited period of time, and probably also in the clinical setting. This does not seem to lead to increased cyclic axial micromotions of the bone fragements. However, the question remains as to how excessive axial displacement of the bone fragment and other loading configurations will affect the process of fracture healing in a patient.
机译:背景和目的:为了减少可能损害骨折愈合的运动,Ilizarov框架中的钢丝预紧力很重要。但是,导线最终将失去部分预张力。为了更深入地了解钢丝预紧的必要性,我们研究了:(1)预紧力损失的数量,(2)预紧力损失与螺栓拧紧扭矩之间的关系,(3)可能导致预紧力丧失的原因;以及(4)预紧力对周期性微动和骨节段总位移的影响。方法:使用MTS机器对3种Ilizarov构型进行了3次测试,这些构型具有各种导线预紧力和螺栓拧紧扭矩。对于每根线材,在整个实验过程中测量其滑移和张力。结果:发现由于打滑导致钢丝预紧力损失(高达75%)。较高的螺栓拧紧扭矩显着减少了预张力损失。此外,较高的线预紧力减小了骨碎片的最大轴向位移。但是,当不存在钢丝预紧力时,轴向轴向运动最少,这可能是由于钢丝没有后坐力。解释:实验性Ilizarov外固定器中的导线会在有限的时间内,甚至在临床环境中失去预张力。这似乎并不会导致骨骼断裂的周期性轴向微动增加。然而,仍然存在关于骨碎片的过度轴向位移和其他载荷构造将如何影响患者的骨折愈合过程的问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号